Seals for Oral-Use Vials and Oral Vaccines

For a vial whose contents are taken by mouth, Autofits makes two seals that come off whole by hand: the FlipTop® Pull Ring, often marked “For Oral Use Only”, at 20 and 32 mm, and the FlipTop® Tear-Down, used mainly for oral vaccines, at 13, 20 and 32 mm. Neither needs a tool, and both leave the full vial mouth open for a dropper or for pouring, where a flip-off seal would leave only a small needle port.
Key takeaways
- Two seals for oral use: FlipTop Pull Ring and FlipTop Tear-Down, both aluminium-plastic combination seals to ISO 8362-6.
- Pull Ring: 20 mm (bridge) and 32 mm (flower), often marked “For Oral Use Only / Flip Top & Tear”.
- Tear-Down: 13, 20 and 32 mm, used mainly for oral vaccinations.
- Opening: both remove the whole seal by hand, turning anticlockwise, with no tool.
- Maximum opening force: 30 N at 13 mm, 35 N at 20 mm, 65 N at 32 mm.
- Delivery forms: Regular, Ready-to-Sterilise and Ready-to-Use.
Why an oral-use vial needs a different seal
A flip-off seal is built to keep the aluminium crimped and expose a small injection point; an oral product needs the whole closure off. Oral vaccines and oral medicines in vials are given with a dropper or poured, so the stopper has to come out and the aluminium has to come away with it. A seal that is removed whole, by hand, is the closure designed for that.
The marking matters as much as the mechanism. Where oral and injectable products of the same neck size are handled in the same clinic or on the same line, a seal that reads “For Oral Use Only” on the closure itself tells the person holding the vial the route of administration before it is opened. Oral vaccines such as the oral polio vaccine described by the World Health Organization are given as drops by mouth, which is the presentation these seals serve.
The two seals compared
| FlipTop® Pull Ring | FlipTop® Tear-Down | |
|---|---|---|
| Main use | Oral medicines, often marked “For Oral Use Only” | Oral vaccinations |
| Sizes | 20 mm (bridge), 32 mm (flower) | 13, 20, 32 mm |
| How it opens | Pull the aluminium ring and turn anticlockwise; the whole seal detaches once the polypropylene disc is detached | Open the polypropylene disc, then pull and rotate it anticlockwise so the whole seal comes off |
| Tool needed | None | None |
| Standard | ISO 8362-6 | ISO 8362-6 |
The two opening actions are compared in detail on pull ring vs tear-down.
Dimensions
From the Autofits engineering drawings, in millimetres.
| Seal | Inner diameter of aluminium | Outer diameter of disc | Total or assembled height | Aluminium thickness | Bridges or fins |
|---|---|---|---|---|---|
| Pull Ring 20 mm | 20.00 to 20.60 | 22.00 to 23.00 | 9.00 to 10.00 | 0.18 to 0.22 | 6 taper bridges |
| Pull Ring 32 mm | 32.15 to 32.40 | 36.20 to 36.60 | 14.30 to 15.30 | 0.20 to 0.26 | 8 fins (flower) |
| Tear-Down 13 mm | 13.30 to 13.50 | 14.80 to 15.40 | 6.80 to 8.30 | 0.18 to 0.22 | Bridge |
| Tear-Down 20 mm | 20.00 to 20.60 | 22.00 to 23.00 | 8.80 to 9.80 | On request | Bridge |
| Tear-Down 32 mm | 32.40 to 33.15 | 35.70 to 36.70 | 14.30 to 15.30 | 0.20 to 0.26 | 8 fins (flower) |
Maximum opening force varies with the number of bridges: 30 N at 13 mm, 35 N at 20 mm and 65 N at 32 mm. Each size has its own page with the full drawing data: Pull Ring 20 mm, Pull Ring 32 mm, Tear-Down 13 mm, Tear-Down 20 mm and Tear-Down 32 mm.
Choosing between them
Choose the Tear-Down for small oral-vaccine vials, and the Pull Ring where the closure must carry the oral-use marking on a 20 or 32 mm vial.
- 13 mm vial, oral vaccine: Tear-Down 13 mm is the only one of the two made at 13 mm.
- 20 mm vial, oral medicine that must be marked: Pull Ring 20 mm.
- 20 mm vial, oral vaccine: either; the Tear-Down is the one most used for vaccinations.
- 32 mm vial: either, in the flower design.
Both are available glossy, matte, embossed or printed, so a Tear-Down can carry an oral-use statement too; finishes and other options are on the custom seals page. Both come in Regular, Ready-to-Sterilise and Ready-to-Use form.
In practice at Autofits
Both seals are made in Nashik on the same lines as the FlipTop Optima: AA8011 aluminium with an epoxy lacquer to 21 CFR 175.300, polypropylene discs moulded in house on Toyo injection machines, assembly in an ISO Class 8 cleanroom and 100% camera inspection, under ISO 15378:2017. How the range serves vaccine makers is covered on the vaccines page.
Frequently asked questions
Which vial seal is used for oral vaccines?
The FlipTop Tear-Down seal is used mainly for oral vaccinations. The polypropylene disc is opened, then pulled and rotated anticlockwise so the whole seal comes off by hand, leaving the vial mouth open for a dropper. It is made in 13, 20 and 32 mm.
What does “For Oral Use Only” on a vial seal mean?
It tells the person handling the vial that the contents are taken by mouth and not injected. Autofits FlipTop Pull Ring seals are often marked “For Oral Use Only / Flip Top & Tear”, so the route of administration is on the closure itself.
How do you open a pull ring vial seal?
Pull the aluminium ring and turn it anticlockwise. Once the polypropylene disc is detached, the whole seal comes away, with no tool needed. Maximum opening force is 35 N at 20 mm and 65 N at 32 mm.
Can an oral-use seal be ready to use?
Yes. FlipTop Pull Ring and Tear-Down seals are available Regular, Ready-to-Sterilise or Ready-to-Use, gamma or ETO sterilised and double-bagged.
What sizes are oral vaccine vial seals made in?
Autofits makes the FlipTop Tear-Down at 13, 20 and 32 mm and the FlipTop Pull Ring at 20 and 32 mm.
Sources
- World Health Organization: Poliomyelitis fact sheet
- ISO: ISO 8362-6, Injection containers and accessories, Caps made of aluminium-plastics combinations (iso.org)
- Autofits FlipTop Seals dossier and engineering drawings
*Last updated: 2026-11-06. Product information only, not regulatory advice.*